ASTM D6978 chemo-rated gloves are the only exam gloves that belong anywhere a hazardous drug is compounded, administered, or cleaned up. The word "chemo" printed on a box is not the rating. The rating is a test result: a glove material held in continuous contact with a panel of chemotherapy drugs, with permeation measured until breakthrough or until 240 minutes run out. If your oncology clinic, infusion center, pharmacy or veterinary practice is buying standard nitrile for that work because it is a few dollars cheaper per case, this is the guide to hand your safety committee.
What ASTM D6978 tests
The standard is ASTM D6978, Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs. ASTM's current listing is D6978-05(2023); the 2019 reapproval is still cited on most glove boxes. The scope describes a protocol that holds glove material in continuous contact with nine chemotherapy drugs under conditions that approximate worst-case clinical use, and reports breakthrough time per drug. ASTM says the data are for relative comparison of glove materials.
The numbers that appear in FDA 510(k) summaries for chemo-rated gloves make the test concrete:
- Breakthrough is detected at a permeation rate of 0.01 micrograms per square centimeter per minute. That is the threshold at which the drug is considered to have come through.
- The maximum test time is 240 minutes. A result of ">240" means the drug did not reach the threshold within the four-hour window. It does not mean the glove is impermeable.
- The drug panel in a typical filing: carmustine, cisplatin, cyclophosphamide, doxorubicin, etoposide, fluorouracil, methotrexate, paclitaxel and thiotepa, each at a stated concentration. Some manufacturers test 25 or more.
That last point matters more than any marketing claim. Two drugs on the panel defeat most glove materials in minutes.
Carmustine and thiotepa: the numbers on the box
Here are the published 510(k) results for three cleared gloves, as filed with FDA:
| Drug (test concentration) | Nitrile exam glove, K110921 | Polyisoprene surgical glove, K110272 | Latex exam glove, K121926 |
|---|---|---|---|
| Carmustine, BCNU (3.3 mg/mL) | 0.61 min | 0.37 min | 15.4 min |
| Thiotepa (10 mg/mL) | 10.6 min | 0.44 min | 1.6 min |
| Cyclophosphamide (20 mg/mL) | >240 min | >240 min | >240 min |
| Doxorubicin HCl (2.0 mg/mL) | >240 min | >240 min | >240 min |
| Fluorouracil (50 mg/mL) | >240 min | >240 min | >240 min |
| Paclitaxel (6.0 mg/mL) | >240 min | >240 min | >240 min |
Every one of those gloves is legitimately chemo-rated. Every one of them is permeated by carmustine in under 16 minutes. The filings say so in plain language: these drugs have "extremely low permeation time of less than 30 minutes." A chemo rating tells you the glove was tested and tells you the breakthrough time for each drug. It is up to your pharmacy and nursing leadership to decide, drug by drug, what glove change interval that implies. Ask your clinical lead or your USP 800 designated person; this is a buying guide, not a handling protocol.
Who needs chemo-rated gloves
USP General Chapter 800 covers the handling of hazardous drugs in healthcare settings. As the Oncology Nursing Society summarized it, the chapter requires double gloving with two pairs of gloves meeting ASTM D6978-05 or its successor. NIOSH, ASHP, ONS and USP all point to D6978 rather than ASTM F739, the general chemical permeation method, because F739's test conditions can produce shorter apparent breakthrough times and a different risk picture.
In purchasing terms, the buyers who should have a D6978-rated glove on contract:
- Hospital and infusion-center pharmacies compounding or repackaging hazardous drugs.
- Oncology nursing and infusion staff administering them.
- Environmental services handling hazardous drug waste and spills in those areas.
- Veterinary oncology practices, which use the same drugs.
- Home infusion and long-term care where hazardous drugs are administered in the room.
Standard nitrile exam gloves tested only to ASTM D6319 are the right glove for everything else. Buying chemo-rated for the whole building is money spent on a rating most of the building does not use.
How to read a glove box
Five lines to find before you approve a chemo-rated glove:
- The claim. "Tested for use with chemotherapy drugs" per ASTM D6978. If the box says "chemo" without the standard, ask for the 510(k) summary.
- The drug table. The IFU or 510(k) summary should list each drug, its concentration, and the breakthrough time. If the manufacturer will not produce it, the rating is a word.
- The 510(k) number. Chemo-rated exam gloves clear under product codes such as LZA and LZC with the chemotherapy claim added. Look it up on FDA's database; the clearance letter names the device.
- Thickness is not the spec. A 4 mil nitrile and a 6 mil nitrile can both pass, and the thicker one is not automatically better against carmustine. The test result is the spec.
- Powder-free and latex status. USP 800 settings double glove, so sizing and donning matter. Latex chemo gloves exist and are cleared; most facilities exclude them for allergy reasons.
The same discipline applies to any glove claim. Our guide on what to check beyond price per box covers 510(k) verification, AQL and origin.
Quote chemo-rated gloves by the test result
Keystone quotes ASTM D6978 chemo-rated nitrile exam gloves and sterile chemo-rated surgical gloves alongside standard D6319 nitrile, with the 510(k) number and the manufacturer's drug table attached to the quote, so your pharmacy signs off on data and not a box color. Quotes go out the same day during business hours, and approved accounts get Net 30. Browse gloves and hand protection or send your hazardous drug areas and glove sizes for a quote.
Sources
- ASTM D6978-05(2019), Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs: https://store.astm.org/d6978-05r19.html
- FDA 510(k) K110921, nitrile exam gloves tested for use with chemotherapy drugs: https://www.accessdata.fda.gov/cdrh_docs/pdf11/K110921.pdf
- FDA 510(k) K110272, sterile polyisoprene surgical gloves tested for use with chemotherapy drugs: https://www.accessdata.fda.gov/cdrh_docs/pdf11/K110272.pdf
- FDA 510(k) K121926, latex exam gloves tested for use with chemotherapy drugs: https://www.accessdata.fda.gov/cdrh_docs/pdf12/K121926.pdf
- Oncology Nursing Society, Handle With Care: How USP 800 Will Affect Nursing Practice (2019): https://www.ons.org/news-and-views/handle-with-care-how-usp-800-will-affect-nursing-practice
- NIOSH (CDC Stacks), Are gloves and gowns safe for handling chemotherapy, Connor et al., 2015: https://stacks.cdc.gov/view/cdc/201032