A recall on a stand-alone product is a shelf pull. A recall on a component sealed inside a custom procedure kit is a different job, and most facilities find that out the week it happens. On October 20, 2026 FDA posted two early alerts for Cardinal Health Presource procedure kits that contain BD Alaris pump infusion sets covered by the BD and CareFusion 303 Class I recall. The kits are not being pulled. The instruction is to find them, quarantine them, label the front of every kit, and have clinicians remove and discard the recalled set before the rest of the kit is used. That is a supply chain task with a clinical deadline, and this is the checklist to run it. Keystone quotes replacement IV administration sets and infusion supplies from BD, B. Braun, ICU Medical and Baxter for licensed facilities the same business day.

What the Alaris set recall covers

The underlying recall started with an FDA early alert on July 15, 2025, was classified Class I on September 29, 2025, and was updated on October 5, 2026 to add more affected products. It covers BD Alaris Pump Infusion Sets used with the Alaris Pump Module model 8100 on Guardrails Suite MX software through version 12.6. The problem is performance at the low end: sets with restrictive components, which FDA lists as closed male luers, filters, check valves, Y-sites and microbore tubing, perform worse than the labeling disclosed. FDA says the deviations most commonly occur at programmed rates under 1 mL per hour and at bolus volumes at or under 5 mL, and that the effects include under-infusion, over-infusion, delayed occlusion alarms and bolus inaccuracy. The patients most exposed are the ones most likely to be on low-rate infusions: neonatal, infant, pediatric and critical care.

The FDA recall database record for the first event lists 33,162,680 units in commerce across 24 catalog numbers, distributed nationwide. As of August 19, 2026, BD had not reported any serious injuries or deaths. FDA's instructions to facilities are specific: immediately identify and discard any set listed in BD's Attachment C of discontinued sets, identify an alternate Alaris pump set that is clinically appropriate for the use case, and if alternates are limited, prioritize them for critical care, neonatal, infant and pediatric areas. Where set performance is not clinically acceptable, FDA points facilities to the Alaris Syringe Module or another syringe pump.

Why the Presource alerts change the job

Cardinal Health builds Presource kits for a single surgical or medical procedure, and some of those kits were packed with the affected Alaris sets. Cardinal sent affected customers a letter on September 24, and FDA's October 20 early alerts restate the instructions:

  • Locate and quarantine affected kits.
  • Alert clinicians of the action.
  • Apply a label to the front of each affected Presource kit instructing clinicians to remove and discard the recalled component.
  • Request additional labels from Cardinal Health if you run short.
  • Distributors that further distributed affected kits must notify their customers.

The second alert adds the use instructions: review BD's customer letter, apply the Attachment B performance disclosures and the Attachment D cautions, and follow the version 12.3 to 12.6 manual mitigations on container height and occlusion clearing. Cardinal's position is that every convenience kit containing an affected set should be corrected, not just the kits on its list. The affected kit numbers are in FDA's "Full List of Affected Devices" attachment on each alert, not in the alert text, so the first task is to download both lists and match them to your item master.

The supply chain checklist

Run this in order. Most of it is work that only materials management can do, because clinicians see a kit label, not a kit bill of materials.

Step Owner Done when
Pull the two FDA affected-kit lists and the BD Attachment C list Supply chain Kit numbers matched to your item master and par locations
Physical count of affected kits in central supply, OR cores, procedure rooms and crash carts Supply chain Count sheet by location, signed
Quarantine or label every affected kit Supply chain Label on the front of every kit, quarantine log updated
Clinician notification with the removal instruction Nursing and OR leadership Sign-off on file
Identify the alternate Alaris set per use case and confirm pump software version Clinical engineering and pharmacy Alternate set numbers approved in writing
Order alternate sets and any kit components lost when the set is removed Supply chain PO confirmed with dates
Return the Cardinal and BD response forms Supply chain Credit or replacement acknowledged
Check every other convenience kit on the shelf for the same set numbers Supply chain Bill of materials review complete

That last row is the one that separates a clean recall from a repeat finding at the next survey. Cardinal has said all convenience kits with an affected set should be corrected. If you buy custom trays from more than one kit maker, pull the bill of materials for each and search it for the Alaris catalog numbers yourself. Do not wait for a second letter.

What removing the set does to the kit

A Presource kit with the set removed is still a sterile kit, but it is no longer complete. The procedure now needs an administration set from another source, delivered to the room, and the circulator has to know that before the case starts. Three options, in order of preference:

  1. Replace the set with the approved alternate Alaris set and stage it with the kit. This keeps the pump workflow unchanged.
  2. Move low-rate infusions to a syringe pump where FDA's performance disclosures make the pump set inappropriate for the drug and rate. This is a clinical decision and belongs to pharmacy and anesthesia.
  3. Re-spec the kit at the next kit review so the set ships separately until BD's replacement sets are in normal supply.

Whichever path you take, add the replacement sets to the par at the same count as the kits on the shelf. A kit count of 200 with a replacement set count of 50 is a stockout in three weeks.

Reading an FDA early alert

FDA's early alert program started as a pilot in November 2024 and was expanded to all medical devices on September 29, 2025. An early alert is FDA's notice of a potentially high-risk device issue, published before the agency has finished classifying the action. That matters for buyers because the alert can land weeks before the formal recall notice, and the facility that acts on the alert has its kits labeled before the distributor's letter arrives. Keystone's supply disruption monitor posts the device recalls and early alerts that touch Keystone's categories, with the FDA link and the date verified, so supply chain sees it the same day.

How Keystone helps

The kit correction itself runs through Cardinal and BD. What Keystone does is close the gap the correction opens: alternate administration sets, extension sets, stopcocks and filters from BD, B. Braun, ICU Medical and Baxter, quoted the same business day with the pump compatibility and the drop factor confirmed, and the kit components you lose when a set comes out. IV sets are prescription products, so quotes go to licensed facilities and substitutions go to your clinical lead for approval before shipping. Approved accounts ship on Net 30 from Edina, Minnesota.

Send the affected set numbers, your kit count and your weekly case volume to the quote desk or call (320) 808-3686. Browse IV and infusion supplies or request a same-day quote, and get the replacement sets on the shelf before the labeled kits run out.